DO Saturation and Sensor Check

A DO analyzer is calibrated against the oxygen that water can hold at saturation, which falls with temperature and elevation. This finds the saturation value for your conditions, shows a reading as percent saturation, and compares an online analyzer with a Winkler or calibrated handheld reading.

DO Saturation and Sensor Check

Formula

DO (fresh water, 1 atm), mg/L = exp[-139.34411 + 1.575701e5/T - 6.642308e7/T² + 1.2438e10/T³ - 8.621949e11/T⁴]
   T = °C + 273.15 (Benson and Krause; Standard Methods 4500-O; USGS)
Salinity factor = exp[-S x (0.017674 - 10.754/T + 2140.7/T²)]
Pressure factor = (P - u)(1 - θP) / [(1 - u)(1 - θ)], P in atm
   u = exp(11.8571 - 3840.70/T - 216961/T²),  θ = 0.000975 - 1.426e-5 t + 6.436e-8 t²
P from elevation: mm Hg = 760 x (1 - 6.8756e-6 x ft)^5.2559
Inputs used
  • Water temperature (°C)
  • Elevation above sea level (ft)
  • Barometric pressure, true (not sea-level corrected) (optional) (mm Hg)
  • Salinity (optional) (ppt (g/kg))
  • Analyzer reading (optional) (mg/L)
  • Reference reading: Winkler or calibrated handheld (optional) (mg/L)
  • Allowed difference (optional) (mg/L)

Worked example (fictional plant data)

  1. Aeration basin at 20 °C, plant elevation 1,000 ft, fresh water
  2. Pressure = 760 x (1 - 6.8756e-6 x 1,000)^5.2559 = 732.9 mm Hg
  3. Saturation at sea level, 20 °C = 9.09 mg/L; corrected for pressure = 8.76 mg/L
  4. Analyzer reads 7.9 mg/L = 90.2% saturation
  5. Winkler reads 8.2 mg/L: difference = -0.30 mg/L, more than the 0.2 mg/L allowed, so recheck the sensor

Cautions

  • Use true (station) barometric pressure for calibration. Weather reports give sea-level corrected pressure, which is wrong for this.
  • Compare readings at the same spot, depth and time. DO changes quickly across an aeration basin.
  • Optical (luminescent) sensors drift less than membrane sensors but still need cap replacement and cleaning.
  • If the analyzer and reference disagree, clean and verify before you recalibrate. Do not calibrate a dirty sensor.

References

  • Standard Methods for the Examination of Water and Wastewater, 4500-O (oxygen solubility table and equations).
  • U.S. Geological Survey, Office of Water Quality Technical Memorandum 2011.03, and the USGS DOTABLES program.
  • Benson, B.B., and Krause, D., 1984, Limnology and Oceanography, v. 29, no. 3, p. 620 to 632.

Results are estimates for operator and technician use. Equations follow the references above; always check against the instrument manual, your calibration procedure and plant data.